Product model | DS-812N | DS-816N |
Measurement condition | d/0(,0)(:CIE No.15,ISO 7724/1,ASTM E1164,DIN 5033 Teil7,JIS Z8722 Condition c,GB/T 3555,ASTM D156,GB/T 6540,ASTM D1500,ASTM D1209,ASTM D5386,ASTM D6045,ASTM D6166,ASTM D1502) | |
Stand-alone operation | Support | |
Integrating sphere | Φ40mm,Avian-D Fully diffuse surface coating | |
Lighting source | CLEDs(Full band balanced Led light source) | |
Inductor | Dual optical sensor array sensor | |
Wavelength range | 360-780nm | |
Wavelength interval | 10nm | |
Half-band width | 5nm | |
Measuring range | 0~200% | |
Resolution | 0.0001 | |
Observer Angle | 2°/10° | |
Observation light source | A,B,C,D50,D55,D65,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,U30,U35,DLF,NBF,TL83,TL84,ID50,ID65,LED-B1,LED-B2,LED-B3,LED-B4,LED-B5,LED-BH1,LED-RGB1,LED-V1,LED-V4 | |
Display | Pharmacopoeia color number, transmission map/data, sample chromaticity value, color difference value/map, pass/fail result, color deviation, color simulation, historical data color simulation, manual input of standard sample, test report | |
Measuring time | ≤1.5s | |
Measuring optical path | Standard: 10mm glass cuvette, 50mm glass cuvette (optional 33mm and 100mm glass cuvette) | |
Color space | CIE Lab,LCh,CIE Luv,XYZ,Yxy,Transmissivity, Hunter Lab Munsell,MI,CMYK | |
Color difference formula | ΔE*ab,ΔE*CH,ΔE*uv,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*94,ΔE*02 | |
Other indicators | Pharmacopoeia (Chinese Pharmacopoeia, European Pharmacopoeia, American Pharmacopoeia),WI(ASTM E313-20,ASTM E313-73,CIE,ISO2470/R457,Hunter,Taube Berger Stensby),YI(ASTM D1925,ASTM E313-20,ASTM E313-73),Tint(ASTM E313-20),Allochromic index Milm,APHA,Hazen,Pt-Co(Platinum-cobalt index),Gardner(Gardner index),Iron cobalt index,Saybolt(Seibert index),ASTM Color,EBC(Beer color),ICUMSA Color(Sugar chroma) | |
Repeatability* |
Transmittance: standard deviation ≤0.08% Colorimetric value: ΔE*ab≤0.015 standard deviation, maximum ≤0.04 |
Transmittance: standard deviation ≤0.08% Chroma value: ΔE*ab≤0.01 standard deviation, maximum ≤0.03 |
Inter instrument agreement | ΔE*ab < 0.25 | ΔE*ab < 0.2 |
Heating function | / | Yes |
Heating temperature range | / | From ambient temperature to 90ºC |
Audit trail | Yes (as required by FDA and CFDA) | |
Special function | Intelligent zero calibration | |
Screen size | 7 inch touch screen | |
Store data | Over 100,000 pieces of data | |
Software support | ColorExpert,Color QC2 |